Datasheet: JDSU-ORL-55 |
 |
| JDSU ORL-55 Smart Optical Return Loss Meter Description: |
 JDSU ORL-55 Smart Optical Return Loss Meter is a high performance instrument for field, production and lab use.

JDSU ORL-55 Smart Optical Return Loss Meter can be used in three different modes as a return loss meter, optical power/loss meter and a laser source.
 |
| JDSU ORL-55 Smart Optical Return Loss Meter Applications: |
| JDSU ORL-55 |
 Designed for fiber optic field use for technicians and Installers/maintenance people within fiber optic networks especially for FTTX, CATV and high bandwidths installations.
 |
| JDSU ORL-55 Smart Optical Return Loss Meter Characteristics: |
| |
 Ensures quick and reliable testing of the return loss of an optical system. High return loss values are essential for ensuring the signal quality especially with CATV & FTTx networks.
 |
| |
 TRIPLEtest function measures return loss at up to three wavelengths simultaneously in real time using digital signal processing.
 |
| |
 Return loss referencing performed via a single key stroke.
 |
| |
 Very easy referencing due to ultra stable lasers and special signal processing.
 |
| |
 Can be used as a standalone optical return loss meter, an optical power/loss meter and a laser source.
 |
| |
 USB interface for easy remote control & report generatio
 |
| |
 Internal memory for over 1000 results
 |
| |
 New Visual Fault Locator Option
 |
| |
 New Host USB data storage option
 |
| |
 Auto-lambda function provides automatic wavelength detection to speed up testing.
 |
| |
 70% reduction in testing time due to multiple automatic functions.
 |
| |
 Industries first auto-zeroing function for very high accuracy with no manual zeroing needed.
 |
| |
 SMARTEnergy power supply management system.
 |
| |
 SMARTStar graphical user interface for fast and easy operation.
 |
| |
 Quick start function requiring no warm up reduces testing time.
 |
| |
 Robust shock proof instrument with great look and feel.
 |
| |
 Built-in real-time clock
 |
| JDSU ORL-55 Smart Optical Return Loss Meter Specifications: |
| Data readout/remote control |
 Via a USB interface
 |
| Modulation detection |
 270 Hz, 1 kHz, 2 kHz
 |
| Display |
 High visibility, 128 x 64 dots with backlight
 |
| Optical connector |
 SM, APC-type
 |
| Adapters |
 Interchangeable for LC, SC, FC, ST, DIN
 |
| Power supply |
 Integrated fast battery charging function (2 hours)
 |
| Four-way powering mechanism |
 AA dry battery, AA NiMH, AC, USB
 |
| Calibration interval |
 3 years
 |
| Ambient temperature |
 Nominal range of use: −10 C to +55 C;
Storage and transport: −40 C to +70 C
 |
| Memory |
 Data memory: 1000 measurement results;
Data readout remote control (via cable K804): client USB interface
USB data storage (option): via host USB interface
 |
| Dimensions |
 W x H x D approximately 95; 60; 195 mm
(3.74; 2.36; 7.68 in)
 |
| Weight |
 Approximately 500 g (1.1 lb)
 |
| Return loss meter: |

 |
| Selectable wavelength options |
 1310/1550 nm;
1310/1490/1550 nm
1310/1550/1625 nm
 |
| Spectral width (RMS) |
 <5 nm
 |
| Display range |
 0 dB to 70 dB
 |
| Measurement range |
 0 dB to 60 dB
 |
| Measurement accuracy |
 +- 0.7 dB (0 dB to 50 dB);
+- 0.9 dB (50 dB to 60 dB)
 |
| Resolution |
 0.01 dB
 |
| Power meter: |

 |
| Wavelength range |
 1260 to 1650 nm
 |
| Factory-calibrated wavelengths |
 1310/1550/1625 nm
 |
| User-calibrated wavelength |
 In 1 nm intervals from 1260 to 1650 nm
 |
| Photo detector |
 Germanium (GE)
 |
| Display modes |
 DB/dBm/W
 |
| Display range |
 −70 to +6 dBm
 |
| Maximum input level |
 +6 dBm
 |
| Resolution |
 0.01 dB, 0.001 μW
 |
| Measurement accuracy |
 +- 0.4 dB
 |
| Laser source: |

 |
| Selectable wavelength options |
 Selectable wavelength options 1310/1550 nm
1310/1490/1550 nm
1310/1550/1625 nm
 |
| Maximum output power |
 −3 dBm
 |
| Adjustable attenuation |
 0 to 7 dB
 |
| Stability |
 +- 0.02 dB
 |
| Operating modes |
 Continuous wave (CW), modulation 270 Hz, 1 kHz, 2 kHz, Auto-lambda (λ)
 |

Need Technical Advise? Please Call: 602-795-4033
Browse Similar Products |