Datasheet: Protek-6030C |
 |
| Protek 6030C 30 MHz Dual Trace Oscilloscope Description: |
 Protek 6030C 30 MHz Dual Trace Oscilloscope functions as a digital oscilloscope LCR Component Tester. Free ground shipping within continental USA.
 |
| Protek 6030C 30 MHz Dual Trace Oscilloscope Characteristics: |
| |
 30 MHz Vertical Bandwidth (-3dB)
 |
| |
 LCR Component Tester
 |
| |
 Alternate Trigger for a stable display of unrelated signals
 |
| |
 X5 Mag for Vertical Input and x10 for Horizontal
 |
| |
 TV Synch Filter
 |
| |
 Trigger Signal Output
 |
| |
 High Brightness CRT with Internal Graticule
 |
| Protek 6030C 30 MHz Dual Trace Oscilloscope Specifications: |
| Vertical Amplifier (CH 1 and CH 2) |
 Sensitivity 5mV-Div to 20V-Div in 12 steps (1-2-5 seq)
1mV-Div to 4V-Div (X5 Mag)
Mag x5
Accuracy ±3 per cent (5 per cent in X5 Mag)
Bandwidth (-3dB) DC- 0Hz to 30MHz
AC- 10Hz to 30MHz
Rise Time < 18ns (<70ns in X5 Mag)
Input Impedance Res- 1MW ±5 per cent, Cap- 30pF
Input Coupling AC, DC, GND
Invert Channel CH2
CH1/CH2 Trigger Output 100mV-Div into 50W
Display Modes CH1, CH2, ALT, Chop, ADD, X-Y
Maximum Input Voltage 400V (DC or AC peak)
 |
| Horizontal Sweep |
 Sweep Time 100ns-Div to 0.2s-Div in 20 steps (1-2-5 seq)
10ns-Div to 20ms-Div (X10 Mag)
Mag x10
Accuracy ±3 per cent (8 per cent X10 Mag)
Sweep Modes TV, Normal and Single
 |
| Triggering |
 Source CH 1, CH 2, ALT, Line and EXT
Slope Positive or Negative adjustable levels
Coupling AC, DC, Normal-TV
Sensitivity (Int) DC to 20MHz 1 Div
Sensitivity (Ext) 0.3V p-p)
Sensitivity (TV Int) 2 Div
Sensitivity (TV Ext) 0.3V p-p
Trigger Lock 50Hz to 10MHz 2 Div
External Trigger Input Impedance- 1MW ±5 per cent
Maximum Input Voltage- 400V (DC or AC p-p)
 |
| Z Axis Modulation |
 Minimum Input Level TTL
Maximum Input Voltage 50V or AC p-p
Polarity Intensity increases as the voltage decreases
Input Resistance 10KW
Bandwidth (-3dB) DC to 5MHz
 |
| X-Y Mode |
 Inputs X-Axis (CH1), Y-Axis (CH2)
Sensitivity Same as CH1
Frequency Response (-3dB) DC- 0Hz to 1MHz
AC- 10Hz to 1MHz
Maximum Voltage Same as CH1
X-Y Phase Difference < 3°
 |
| Calibrator Signal |
 Waveform Square Wave;
Amplitude 0.5V +- 2 percent p-p;
Frequency 1KHz +- 2 per cent
 |
| Component Test |
 A 60Hz, 9V RMS signal is applied to the CH1 and CH2 input when the AC, DC, GND is in the GND position and the component test button is pushed in.
 |
| General Specifications |
 CRT 8cm x 10cm (3.15" x 3.94")
Power Requirements AC Input Voltage- 120V or 220V
Frequency 50-60Hz
Power Consumption 35W
Operating Temperature 0 to 40°C (32 to 104°F)
Size 5.1" (H) x 12.2" (W) x 16.45" (D)
Weight 14.3 lbs
 |
| Protek 6030C 30 MHz Dual Trace Oscilloscope Accessories Optional: |
| Accessories |
 Manual, Line cord, X1X10 Probes (2), Test Leads
 |